恒溫水浴振蕩器常見故障及處理
恒(heng)溫(wen)水浴振蕩器具有結(jie)構合理、操(cao)作(zuo)簡便、穩定性(xing)能高(gao)等特點(dian),是實(shi)驗室工(gong)作(zuo)人員(yuan)得心應手的理想設備(bei)。
恒溫水浴振蕩器常見故障及處理
1、故障現(xian)象:恒(heng)溫水浴振蕩(dang)器加熱正常,振蕩(dang)無法工作。
解決方法:① 打開控制柜側板,接通電源,將溫控設(she)定為(wei)0度(即(ji)(ji)不加熱狀態),打開振蕩開關(guan),將速度調至(zhi)(zhi)最高;② 用(yong)萬用(yong)表(交流(liu)250檔(dang))測量(liang)變壓(ya)(ya)器(qi)有(you)(you)無220V電壓(ya)(ya)輸入,若無則開關(guan)壞(huai)(huai),反之,測量(liang)變壓(ya)(ya)器(qi)有(you)(you)無12V電壓(ya)(ya)輸出(chu)(用(yong)交流(liu)50V檔(dang)),無則變壓(ya)(ya)器(qi)壞(huai)(huai);③ 此時切勿輕易更(geng)(geng)換(huan)變壓(ya)(ya)器(qi),該情(qing)(qing)況多為(wei)線路板中整(zheng)流(liu)部(bu)分或(huo)電機出(chu)現短(duan)路故障(出(chu)現此種現象,變壓(ya)(ya)器(qi)外觀有(you)(you)燒焦現象),在排除以上(shang)情(qing)(qing)況,更(geng)(geng)換(huan)即(ji)(ji)可;④ 12V電壓(ya)(ya)正(zheng)常,檢(jian)查(cha)有(you)(you)無直(zhi)流(liu)12電壓(ya)(ya)(直(zhi)流(liu)50V檔(dang)),沒(mei)有(you)(you)則整(zheng)流(liu)橋斷路,更(geng)(geng)換(huan);⑤ 有(you)(you)則檢(jian)查(cha)有(you)(you)無電壓(ya)(ya)輸出(chu)至(zhi)(zhi)電機(5-8V左右),有(you)(you)則電機壞(huai)(huai)或(huo)由于儀器(qi)長(chang)時間(jian)沒(mei)有(you)(you)使(shi)用(yong),機械部(bu)分銹死,沒(mei)有(you)(you)則控制板中大(da)功率管壞(huai)(huai)。
2、故障現(xian)象:加熱正常,振蕩(dang)速度變(bian)慢(man)。
解(jie)決方(fang)法:打開控制柜(ju)側板,接(jie)通(tong)電(dian)源,將溫控設定為0度(du)(du)(即不加熱狀(zhuang)態),打開振蕩開關,將速(su)度(du)(du)調至最高,找(zhao)到速(su)度(du)(du)控制板上的兩個(ge)可(ke)(ke)調電(dian)阻,逆時針調整右邊的即可(ke)(ke)解(jie)決。
3、故障現象:恒溫水浴振蕩器(qi)不加(jia)熱,振蕩正(zheng)常。
1)顯示正常,設定正常,無法(fa)加熱(re)(re)。該故障解決方(fang)法(fa):接通電源,打開電源開關,調整設定溫(wen)度高于實際測量(liang)溫(wen)度,檢查溫(wen)控儀有(you)(you)(you)無輸(shu)出指示,有(you)(you)(you)則測量(liang)加熱(re)(re)管是否有(you)(you)(you)電壓輸(shu)入,有(you)(you)(you)則加熱(re)(re)管壞(huai),更換即可,沒(mei)(mei)有(you)(you)(you)電壓輸(shu)入加熱(re)(re)管,多為(wei)繼電器發生故障。若溫(wen)控儀沒(mei)(mei)有(you)(you)(you)電壓輸(shu)出指示,建議(yi)更換溫(wen)控儀。
2)溫控(kong)儀顯(xian)示LLL或999(根(gen)據廠(chang)家(jia)不(bu)同,參數有所(suo)不(bu)同)該故障為(wei)溫度傳(chuan)感(gan)器(qi)斷路(lu)或短路(lu),更(geng)換傳(chuan)感(gan)器(qi)即可。
3)溫度(du)不受控制繼電器壞(huai)或傳感器受潮。