恒溫水浴振蕩器具有結構合理、操作簡便(bian)、穩定性能(neng)高等特點,是(shi)實驗室工作人(ren)員得心應(ying)手的理想設備。
恒溫水浴振蕩器常見故障及處理
1、故障現象:恒(heng)溫水浴振蕩器加熱正常,振蕩無法工作。
解決方法:① 打開控制(zhi)柜側板,接通電(dian)(dian)源,將(jiang)溫控設(she)定(ding)為0度(即不加熱狀態),打開振蕩(dang)開關(guan),將(jiang)速度調至最高;② 用(yong)萬用(yong)表(交(jiao)流250檔(dang))測量(liang)(liang)變壓(ya)(ya)(ya)器(qi)有(you)(you)無(wu)(wu)220V電(dian)(dian)壓(ya)(ya)(ya)輸入(ru),若(ruo)無(wu)(wu)則(ze)(ze)開關(guan)壞(huai)(huai),反之,測量(liang)(liang)變壓(ya)(ya)(ya)器(qi)有(you)(you)無(wu)(wu)12V電(dian)(dian)壓(ya)(ya)(ya)輸出(chu)(用(yong)交(jiao)流50V檔(dang)),無(wu)(wu)則(ze)(ze)變壓(ya)(ya)(ya)器(qi)壞(huai)(huai);③ 此時切勿輕易更(geng)換變壓(ya)(ya)(ya)器(qi),該情(qing)況(kuang)多(duo)為線(xian)路(lu)板中整流部分或(huo)電(dian)(dian)機出(chu)現(xian)短路(lu)故障(出(chu)現(xian)此種現(xian)象(xiang),變壓(ya)(ya)(ya)器(qi)外觀有(you)(you)燒焦現(xian)象(xiang)),在排(pai)除以(yi)上情(qing)況(kuang),更(geng)換即可;④ 12V電(dian)(dian)壓(ya)(ya)(ya)正常,檢(jian)查(cha)有(you)(you)無(wu)(wu)直流12電(dian)(dian)壓(ya)(ya)(ya)(直流50V檔(dang)),沒有(you)(you)則(ze)(ze)整流橋斷(duan)路(lu),更(geng)換;⑤ 有(you)(you)則(ze)(ze)檢(jian)查(cha)有(you)(you)無(wu)(wu)電(dian)(dian)壓(ya)(ya)(ya)輸出(chu)至電(dian)(dian)機(5-8V左(zuo)右),有(you)(you)則(ze)(ze)電(dian)(dian)機壞(huai)(huai)或(huo)由于儀器(qi)長時間沒有(you)(you)使(shi)用(yong),機械部分銹死,沒有(you)(you)則(ze)(ze)控制(zhi)板中大功率管(guan)壞(huai)(huai)。
2、故障現象:加熱正常,振蕩速(su)度變慢。
解(jie)決(jue)方法:打開(kai)控制柜側(ce)板(ban),接(jie)通電源,將(jiang)溫控設定為0度(即(ji)不加熱狀態),打開(kai)振蕩開(kai)關,將(jiang)速(su)度調至最高,找到(dao)速(su)度控制板(ban)上(shang)的兩個可調電阻(zu),逆時(shi)針調整右(you)邊的即(ji)可解(jie)決(jue)。
3、故障現象:恒溫水(shui)浴振蕩(dang)器(qi)不加熱,振蕩(dang)正常。
1)顯示(shi)正常,設定正常,無法加(jia)熱(re)。該故障解決方法:接通電源,打開電源開關,調整設定溫度高于實際測量(liang)溫度,檢查溫控(kong)儀有(you)無輸出(chu)指示(shi),有(you)則(ze)測量(liang)加(jia)熱(re)管(guan)是否(fou)有(you)電壓(ya)輸入,有(you)則(ze)加(jia)熱(re)管(guan)壞,更換(huan)即可,沒有(you)電壓(ya)輸入加(jia)熱(re)管(guan),多為繼電器發生故障。若溫控(kong)儀沒有(you)電壓(ya)輸出(chu)指示(shi),建(jian)議更換(huan)溫控(kong)儀。
2)溫控儀(yi)顯示LLL或(huo)999(根據廠家不(bu)同(tong),參數有所(suo)不(bu)同(tong))該故障為溫度傳(chuan)(chuan)感器斷路(lu)或(huo)短路(lu),更換(huan)傳(chuan)(chuan)感器即可。
3)溫度不(bu)受控制繼(ji)電器壞或傳感器受潮。